Measuring Grip Strength

From Bridges Lab Protocols
Revision as of 17:20, 14 February 2014 by Davebrid (talk | contribs) (wrote grip protocol)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.

Materials

  • Grip Strength Meter: Use the Metek DFE Grip Meter from the Neuroscience Institute Behavioral Core
  • Mice: You can either do it downstairs or bring the mice upstairs. If you move the mice then make sure they have some time to settle first.
  • Prepare a data sheet with one row per animal and space for 3-5 measurements per animal.

Protocol

  • Make sure the meter is charged or plugged in and sitting stably on the bench
  • Turn it on, click on the top left button (MODE) until the measurement display says T-PK (Peak Tension) not NORMAL or C-PK (Compression Peak)
  • Zero the reading by pressing ZERO, the reading should be at or near 0.00 N (measure force in Newtons unless a compelling reason to do this otherwise)
  • Pick up the mouse from the cage, and determine its ear tag number.
  • Place the mice on the grate so that is securely grips with all four paws, holding it by the tail.
  • Gently pull the mouse horizontally by the tail away from the sensor until the mouse lets go.
  • Record the peak grip on your data sheet
  • Let the mouse recover for 2-3s, then repeat 3-5 times in total
  • Move on to the next mouse