BHB Assay

From Bridges Lab Protocols
Jump to navigation Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.

Materials

  • Wako Autokit Total Ketone Bodies: Cat#'s 415-73301, 411-73401 and 412-73791. Once R1 and R2 are resuspended, use within 3 weeks. Protocol modified from Wako Kit Protocol
  • Serum, prepared according to Measuring Fasting Insulin
  • Microtiter plate
  • Plate reader, warmed to 37C

Procedure

  1. Prepare R1 and R2 if not already resuspended
  2. Pipet 4 uL of sample into each well (For ketogenic diet samples pipet 1 uL).
  3. Pipet 0, 1, 2, 3, 4, 5, 6, and 7 uL of 40 uM standard into wells. If evaluating ketogenic diet samples, also pipet 1, 2, 3, 4, 5, 6, 7 and 8 uL of 300 uM standard
  4. Using multichannel pipet add 270 uL of R1 to each well
  5. Incubate at 37C for 5 minutes
  6. Add 90 uL of R2 to each well and mix
  7. Put in plate reader at 37C for 2 minutes
  8. Measure A405 at 30s intervals. You may need to increase interval time in order to measure all your wells. Measure for at least 3 minutes total. Modify BHB Assay.prt on the plate reader if necessary.
  9. Calculate Max V according to the software.
  10. Calculate Ketone Levels by plotting Rate (Max V) versus BHB amount and adjusting by volume used.