Materials

  • FujiFilm Autokit Glucose (Cat 997-03001). Dissolve the Color Reagent in Buffer Solution (150 mL) to make a bottle of 150 mL Color Solution. The solution is stable for 2 weeks at 4C.
  • Glucose Standard Solution (200 mg/dL)
  • Serum samples (thaw on ice if from -80C)

Procedure

  • Plan out 96-well plate, including samples and space for 5 standards and a blank, in duplicates.
  • Add 100 uL of Color Solution to each well.
  • Add 0, 1, 2, 3, 4, 5 uL to each of the standard wells.
  • Add 3 uL of sample to each well.
  • Incubate at room temperature for 30 mins (or 5 mins at 37C).
  • Read plate at 505 nM and 600 nM.

Calculations

  • Use the standard curve to calculate glucose levels.