Difference between revisions of "French Press"
From Bridges Lab Protocols
Davebridges (Talk | contribs) (imported protocol) |
(No difference)
|
Revision as of 12:22, 4 May 2009
Prepare the Cell (Not Auto-Fill)
- Invert cell on black stand without base piece. Remove piston to fill line
- Pour in sample, replace base and place back on the press. Ensure valves are tightened
Loading Sample
- With the back valve open and the bottom valve closed insert tubing into sample
- Set French Press to Down and start. Suction caused by the piston moving out of the cell will bring sample into the cell
- When the base is at the bottom close the intake valve and stop the Press
Lysing Sample
- For bacteria 15 000 psi of cell pressure is sufficient. For a 1cm cell this corresponds to about 1100 psi on the gauge (see chart)
- Turn black knob all the way to the left
- Set Press to High and ensure both valves are securely closed
- Start the press. The cell will move until it encounters pressure and will stop moving (the motor will still be running)
- Slowly turn the knob until the gauge pressure reads desired pressure
- Slowly open valve until sample comes out, keep an eye on the pressure and make sure it doesn’t drop. Continue until piston is fully depressed. Close bottom valve
- Reload sample if necessary
Cleaning/Storage
- Disassemble and rinse all parts with 70% ethanol then distilled water
- Normally cells are stored at 4C to reduce sample heating