== Materials ==
* FujiFilm Autokit Glucose (Cat 997-03001). Dissolve the Color Reagent in Buffer Solution (150 mL) to make a bottle of 150 mL Color Solution. The solution is stable for 2 weeks at 4C.
* Glucose Standard Solution (200 mg/dL)
* Serum samples (thaw on ice if from -80C)
== Procedure ==
== Procedure ==
* Plan out 96-well plate, including samples and space for 5 standards and a blank, in duplicates.  
* Plan out 96-well plate, including samples and space for 5 standards and a blank, in duplicates.  
Line 5: Line 10:
* Add 3 uL of sample to each well.  
* Add 3 uL of sample to each well.  
* Incubate at room temperature for 30 mins (or 5 mins at 37C).  
* Incubate at room temperature for 30 mins (or 5 mins at 37C).  
* Read plate at 505 nM and 600 nM.
* Read plate at 505 nM and 600 nM.
 
== Calculations ==
* Use the standard curve to calculate glucose levels.

Revision as of 00:51, 13 February 2026

Materials

  • FujiFilm Autokit Glucose (Cat 997-03001). Dissolve the Color Reagent in Buffer Solution (150 mL) to make a bottle of 150 mL Color Solution. The solution is stable for 2 weeks at 4C.
  • Glucose Standard Solution (200 mg/dL)
  • Serum samples (thaw on ice if from -80C)

Procedure

  • Plan out 96-well plate, including samples and space for 5 standards and a blank, in duplicates.
  • Add 100 uL of Color Solution to each well.
  • Add 0, 1, 2, 3, 4, 5 uL to each of the standard wells.
  • Add 3 uL of sample to each well.
  • Incubate at room temperature for 30 mins (or 5 mins at 37C).
  • Read plate at 505 nM and 600 nM.

Calculations

  • Use the standard curve to calculate glucose levels.