Glucose Assay in Serum: Difference between revisions
From Bridges Lab Protocols
Jump to navigationJump to search
| Line 21: | Line 21: | ||
* Set ‘Wavelengths’ to “2”. | * Set ‘Wavelengths’ to “2”. | ||
* Set ‘Lm1’ to “490”, ‘Lm2’ to “650”. | * Set ‘Lm1’ to “490”, ‘Lm2’ to “650”. | ||
* Put plate in, REMOVE THE LID. | * Put plate in, '''REMOVE THE LID'''. | ||
* Click ‘Read’. | * Click ‘Read’. | ||
* Reading: top numbers are 490 uL, bottle numbers are 650 uL. | * Reading: top numbers are 490 uL, bottle numbers are 650 uL. | ||
* To export files: click ‘file’ - ‘export’ - ‘save to computer’ - ‘Bridges’ folder - place in USB. | * To export files: click ‘file’ - ‘export’ - ‘save to computer’ - ‘Bridges’ folder - place in USB. | ||
Latest revision as of 00:55, 13 February 2026
Materials
- FujiFilm Autokit Glucose (Cat 997-03001). Dissolve the Color Reagent in Buffer Solution (150 mL) to make a bottle of 150 mL Color Solution. The solution is stable for 2 weeks at 4C.
- Glucose Standard Solution (200 mg/dL)
- Serum samples (thaw on ice if from -80C)
Procedure
- Plan out 96-well plate, including samples and space for 5 standards and a blank, in duplicates.
- Add 100 uL of Color Solution to each well.
- Add 0, 1, 2, 3, 4, 5 uL to each of the standard wells.
- Add 3 uL of sample to each well.
- Incubate at room temperature for 30 mins (or 5 mins at 37C).
- Read plate at 505 nM and 600 nM.
Calculations
- Use the standard curve to calculate glucose levels.
Reading Plate Instructions
- Turn on the plate reader (switch at the back) and the computer.
- Open ‘Soft Max Pro’.
- Click ‘Settings’.
- Set ‘Wavelengths’ to “2”.
- Set ‘Lm1’ to “490”, ‘Lm2’ to “650”.
- Put plate in, REMOVE THE LID.
- Click ‘Read’.
- Reading: top numbers are 490 uL, bottle numbers are 650 uL.
- To export files: click ‘file’ - ‘export’ - ‘save to computer’ - ‘Bridges’ folder - place in USB.