Seahorse XF Media: Difference between revisions

Wrote initial page
 
updated stock concentrations
Line 1: Line 1:
This minimal media stock is the base media for both the glycolysis and mitochondrial stress assays.  It contains no buffering and no carbon sources.  It is a non-buffered media but still needs to be brought to pH 7.4 and checked before each assay.
This minimal media stock is the base media for both the glycolysis and mitochondrial stress assays.  It contains no buffering and no carbon sources.  It is a non-buffered media but still needs to be brought to pH 7.4 and checked before each assay.


For glycolysis assays add:
For glycolysis assays add to warmed XF base media just before the assay and recheck pH:


* 1-4 mM L-Glutamine
* 1-4 mM L-Glutamine (Generally 2 mM is 1mL/100mL of 200 mM stock)


For mitochondrial assays add:
For mitochondrial assays and cellular phenotype assays add to warmed XF base media just before the assay and recheck pH:


* 0.5-10 mM Sodium pyruvate
* 0.5-10 mM Sodium pyruvate (Generally 1 mM is 100uL/100mL of 1M stock)
* 1-4 mM L-Glutamine
* 1-4 mM L-Glutamine (Generally 2 mM is 1mL/100mL of 200 mM stock)
* 2-5 mM Glucose
* 2-25 mM Glucose (Generally 10 mM is 1mL/100mL of 1M stock)


To make base media add the following:
To make base media add the following: